Temperature Measurement Method by Infrared Radiation Pyrometer Using Chopped Broadband Radiation - Signal Conditioning and Acquisition Elements Design
DOI:
https://doi.org/10.31695/IJASRE.2020.33936Keywords:
Contactless temperature sensor, Pyrometer, Infrared, Chopped Broadband, Signal Conditioning, Detector, Lock-In AmplifierAbstract
In this paper, we report the design of a signal conditioning and acquisition elements of a chopped broadband radiation pyrometer for measuring temperature in the range of 900 0C to 1200 0 C. We also report the selection of a suitable chopping frequency within the physical properties of the system and provided the calibration characteristics for the pyrometer. Based on the results, it was found that the designed signal conditioning circuit was able to produce an output voltage of 0.24 – 0.60 V corresponding to the temperature measuring range of the pyrometer. This was achieved through the use of a phase-sensitive detector (PSD) in the form of a lock-in amplifier in combination with a low pass filter. A suitable chopping frequency of 30 Hz was chosen for the shutter pulse so that the detector can have sufficient time to settle before the next shutter pulse transition. Finally, the result also shows that the calibration characteristic of the pyrometer presumes an approximately linear relationship.
References
Bentley, J. P. 2005. Principles of Measurement Systems, Fourth Edition, Prentice Hall, USA. ISBN: 978 – 0 – 13043028 – 1.
Jack, E. K., Etu, I., and Ukanide, N. V. 2016. The Design of Signal Conditioning and Acquisition Elements of Chopped Broadband Radiation Pyrometer. In Proceeding of the World Congress on Engineering and Computer Science, Vol. 1.
Alan, M. S. 2001. Measurement and Instrumentation Principles, Third Edition, Butterworth-Heinemann, Uk.
Bieleck, Z. and Chrzanowski, K. 1999. Infrared Pyrometer for Temperature Measurement of Objects for both Wavelength and Time Dependent Emissivity. Optica Applicata, Vol. XXIX (3), Pp. 285 – 292.
Muller, B., and Renz, U. 2001. Development of Fast Fiber – Optic Two Color Pyrometer for the Temperature Measurement of Surfaces with Varying Emissivities. Review of Scientific Instruments, Vol. 72 (8), Pp. 3366 – 3374.
Ng, D., and Fralick, G. 2001. Use a Multiwavelength Pyrometer in Several Elevated Temperature Aerospace applications. Review of Scientific Instruments, Vol. 72 (1), Pp. 1522 – 1530.
Ueda, T., Hosokawa, A., and Yamamoto, A. 1985. Studies on Temperature of Abrasive Grains in Grinding – Application of Infrared Radiation Pyrometer. Journal of Engineering for Industry, Transactions of the ASME, Vol. 107, Pp. 127 – 133.
Vàzquez, C., Pérez-Prieto, S., López-Cardona, D. J., Tapetado, A., Blanco, E., Moreno-López, J., Montero, S. D. and Lallana, C. P. 2018. Fiber-Optic Pyrometer with Optically Powered Switch for Temperature Measurement. Sensors, Vol. 18 (483), Pp. 1 – 10.
Wallace, P. 2010. Instrumentation for Mechanical Condition Monitoring: Module Notes
Downloads
How to Cite
Issue
Section
License
Copyright (c) 2020 K. T. Aminu, D. D. Yakubu, S.A. Baraza, B. D. Halilu

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.